CLARITASNucMed

Vendor-neutral software · PET · PET-CT · PET-MRI

A number a
radiologist can
defend.

Claritas NucMed builds enhancement, automated segmentation and quantification software for nuclear medicine. Thresholds are set by the clinician. Values are computed at voxel level. Nothing is added to the image that the acquisition did not contain.

ISO 13485 certified · Portfolio cleared by FDA, TGA, ANVISA, CDSCO and HSA · Vendor-neutral · Isotope-agnostic

Axial · illustrative phantom · not patient data SUV 16 8 0

Segmentation threshold

Live
4.0SUVbw min
1.07.514.0
Regions above threshold
Total segmented volume
cm³
Highest SUVmax
Total lesion glycolysis

Illustrative phantom for demonstration. Not patient data and not a diagnostic display. In use, iPETcertum computes these values from the loaded study at voxel level against clinician-defined minimum and maximum SUV and volume parameters.

01

Products

Three cleared software devices spanning acquisition, reading and quantification in nuclear medicine. All brand-agnostic, all deployed into existing scanner and PACS environments rather than alongside them, all developed under an ISO 13485 quality system.

iPET

Enhancement & reconstruction

Noise suppression and reconstruction that recovers diagnostic quality from shorter acquisitions and lower administered activity. The original scan is preserved unaltered; only noise is suppressed and detail already captured is enhanced.

  • Fine lesions retained, no artefacts introduced
  • SUV statistics reported per metric, not as one number
  • Reported against RMSE and SNR
  • Works on new and older installed equipment
  • PET, PET-CT and PET-MRI

iPETcertum

Segmentation & quantification

Segments regions of interest against clinician-defined SUV and volume parameters, and quantifies each by uptake and volume at voxel level. Runs on the original scan and on the iPET-enhanced scan, so the two can be compared directly.

  • Single-click ROI visualisation
  • SUV and volume readout — whole body or selected anatomy
  • Agnostic to radiopharmaceutical and scanner
  • Reproducible between readers
Whole-body PET maximum intensity projection with three regions of interest segmented and colour-coded.

Three regions returned above threshold, each colour-coded and quantified for volume, SUV and SUVmax.

Whole-body PET maximum intensity projection with segmented regions of interest in the abdomen.

Segmentation follows the clinician's SUV and volume range — physiological uptake outside those bounds is not returned.

iPETcertum on whole-body PET. The reader sets minimum and maximum SUV and lesion volume; every region meeting those criteria is segmented and measured in a single pass. De-identified studies from clinical evaluation.

neuroCloud

Brain quantification

Detection, localisation and quantification across the modalities used in neurological assessment — including fused SPECT-MRI subtraction — referenced to normal population data, so a regional value carries a comparison and not only a measurement.

  • PET — hypo- and hypermetabolism, amyloid
  • MRI — volumetric atrophy and lesion analysis
  • SPECT — quantitative DaT analysis
02

Regulatory status

Stated per product and per jurisdiction, because that is the only form in which the question can be answered honestly. Clearance in one market is not clearance in another, and we do not aggregate the two.

Current as at August 2026 — full regulatory annex on request

iPET

PET · PET-CT · PET-MRI

  • United StatesFDA 510(k) K213140
  • BrazilANVISA
  • SingaporeHSA
  • AustraliaTGA
  • IndiaCDSCO 000580
  • EuropeMDR — in progress

iPETcertum

PET · PET-CT · PET-MRI

  • United StatesFDA 510(k) K244016
  • BrazilANVISA 1252201/24-9
  • SingaporeHSA — in progress
  • AustraliaTGA — in progress
  • IndiaCDSCO — in progress
  • EuropeMDR — in progress

neuroCloud

PET · MRI · SPECT

  • BrazilANVISA
  • United StatesFDA — in progress
  • SingaporeHSA — in progress
  • AustraliaTGA — in progress
  • IndiaCDSCO — in progress
  • EuropeMDR — in progress

Cleared and available for sale   Under review   ·   Jurisdictions not listed have no current application.

QUALITY SYSTEM  ·  ISO 13485 certified — DE0507340
VALIDATION  ·  Clinical evaluation conducted across sites in the UK, Asia, India and the Americas
DATA  ·  HIPAA, GDPR and DPDPA compliant; residency configurable by deployment

Performance figures for scan-time and dose reduction are protocol-, scanner- and isotope-dependent. Evaluation data and the conditions under which the figures were obtained are available to qualified partners on request. Claritas does not publish headline performance claims without the acquisition conditions attached.

03

What changes on a scan day

Two constraints govern a nuclear medicine department: how long a patient occupies the scanner, and how long a physician spends drawing round lesions. Claritas works on both, and leaves the diagnosis where it belongs.

Without

  • Full-length acquisition per patient, fixed by protocol and by how noisy the scanner is.
  • Administered activity set high enough to overcome that noise.
  • The physician contours regions by hand — slow, expertise-dependent, and the point at which small or low-uptake lesions get missed.
  • Two readers measuring the same study can report different numbers.
  • Older installed scanners produce images that limit what can be reported from them.

With Claritas

  • Diagnostic quality recovered from shorter acquisitions, so the bed turns over faster.
  • Lower administered activity for the same read — less exposure, less isotope purchased per study.
  • Regions segmented automatically against the physician's own SUV and volume criteria, in one click.
  • The same criteria applied to the same study return the same values, whoever is reading.
  • Installed equipment keeps producing reportable studies without capital replacement.

The department

Scanner time is the binding constraint on a PET service, and it is capital. Shorter acquisitions raise the number of studies a scanner completes in a day without adding a scanner. Reduced administered activity lowers isotope spend on every study — a recurring cost, not a one-off.

The physician

Manual segmentation is labour, not judgement. Removing it returns the time to diagnosis and treatment decisions, and removes the step where a small or faintly avid lesion is most likely to be passed over. The reader still sets the criteria and still makes the call.

The patient

Less time on the table, which matters most to the patients least able to tolerate it, and a lower administered dose for the same diagnostic answer.

The trial sponsor

Response assessment depends on measurements being comparable between timepoints and between sites. Fixed criteria applied identically across a study remove reader drift as a source of variance in the endpoint.

Measured, with the conditions attached

50%+50%Scan time and isotope dose reduced simultaneously, retaining diagnostic quality
80%Maximum acquisition time reduction achieved in clinical evaluation
≥50%Minimum administered isotope dose reduction achieved
<1%SUVmin difference between original and enhanced scan — the noise floor is suppressed without shifting the baseline
Whole-body PET reconstructed from 5 percent of the acquisition, heavily degraded by noise.

5% acquisitionunprocessed

RMSE 219  ·  SNR 10

The same 5 percent acquisition after iPET processing, with organ boundaries recovered.

5% acquisitioniPET processed

RMSE 102  ·  SNR 49

Whole-body PET reconstructed from 20 percent of the acquisition.

20% acquisitionunprocessed

RMSE 109  ·  SNR 40

The same 20 percent acquisition after iPET processing.

20% acquisitioniPET processed

RMSE 71  ·  SNR 95

The same patient, reconstructed from truncated acquisitions. Read left to right: a 5% acquisition processed by iPET carries less error and more signal than a 20% acquisition left alone.
RMSE · LOWER IS BETTER 219 102 5% scan time 109 71 20% scan time SNR · HIGHER IS BETTER 10 49 5% scan time 40 95 20% scan time Reduced acquisition, unprocessed Same acquisition, iPET processed
Reconstruction quality recovered from truncated acquisitions. A 5% acquisition processed by iPET reaches a lower error and a higher signal-to-noise ratio than an unprocessed 20% acquisition.

Different SUV statistics behave differently under enhancement, and we report them separately rather than as a single number. SUVmin is preserved to within 1% — the noise floor falls without the baseline moving. SUVmean over a defined region is substantially preserved. SUVmax is an extreme-value statistic and is expected to move, because in a short, noisy acquisition the hottest voxel is hot partly because of noise; suppressing that noise brings the value closer to the long-acquisition reference, not further from it. Gains are protocol-, scanner- and isotope-dependent: up to 4–5× SNR improvement and 50% RMSE reduction on low-dose or reduced-time acquisitions, around 20% SNR and 10% RMSE on full-dose studies, and performance holds for high-BMI patients. Full evaluation data, with the acquisition conditions and smoothing parameters under which each figure was obtained, is available to qualified partners.

04

neuroCloud

Brain quantification across the three modalities used in neurological assessment. Detection, localisation and measurement, referenced to normal population data — so a regional value arrives with a comparison attached, not just a number.

PET

Metabolic and amyloid analysis. Regional uptake against age-matched normative data, and amyloid burden on the Centiloid scale.

MRI

Volumetric analysis and lesion segmentation. Whole-brain and tissue-class volumes with percentile placement.

SPECT

Quantitative dopamine transporter analysis by striatal subregion, with normative referencing and serial comparison.

Longitudinal lesion analysis MRI

Two studies of the same patient, acquired a year apart, coregistered and compared. Every lesion is classified by anatomical class under the McDonald criteria and by what it did between scans — unchanged, smaller, larger, or new. The clinician sees the map; the report carries the arithmetic.

BaselineFollow-upChange map
Three rows of matched axial FLAIR slices: baseline lesion segmentation, follow-up segmentation, and the change map between them.
Matched axial slices from two timepoints. The third row is the difference — not a second read, a computed comparison. De-identified study, reproduced with permission.

Unchanged   Smaller   Larger   New

Axial slices with lesions colour-coded by change category between the two timepoints.

Each lesion is colour-coded by what it did between studies, in axial, coronal and sagittal reconstruction.

Lesion load by class — illustrative case
ClassBaselineFollow-upΔ vol.
Periventricular21 · 3.4319 · 4.64+35.3%
Deep white matter13 · 0.7319 · 0.96+31.5%
Infratentorial0 · 0.001 · 0.03new
Total34 · 4.1639 · 5.63+35.3%

Counts and volume in cm³. Note the periventricular row: two fewer lesions, 35% more lesion volume. Existing lesions grew and merged. A count alone would have read as improvement.

Global volumetry with normative percentile
MetricVolumePctl.Annual rate
Grey matter618.5345−1.56%
White matter540.9352−1.33%
Brain parenchymal fraction77.7112−0.44%

Volume in cm³, percentile against the normal population. The annual rate is the output that matters clinically — it turns two scans into a trajectory.

Regional metabolic analysis PET

FDG PET coregistered to CT, parcellated to an anatomical atlas, and compared region by region against an age-matched normative population. Every region returns a mean value, a percentage deviation, a z-score and a percentile — left and right separately, with the asymmetry computed between them.

2.0–2.5 SD   2.5–3.0 SD   3.0–6.0 SD

Axial FDG PET slices fused with CT, with hypometabolic regions shaded by standard deviation from the normative mean.

Regions falling below the normative range are shaded by severity, not simply flagged. The reader sees how far outside normal a region sits, and where the boundary lies.

Five most significant regions — illustrative case
RegionΔ%ZPctl.
Superior frontal gyrus L−11.40−2.880.00
Middle frontal gyrus L−6.66−2.153.23
Broca's area L−7.94−1.793.23
Superior frontal gyrus R−6.16−1.714.84
Anterior orbital gyrus L−6.83−1.643.23

The top row is the finding: a left superior frontal gyrus at the 0.00 percentile — below every subject in the normative cohort. Its right-side counterpart sits at 4.84, giving a measured asymmetry rather than an impression of one.

Normative comparison curves showing the patient value plotted against age-referenced percentile bands for five regions.

Each region is plotted against the age-referenced normative distribution, so the comparison is visible rather than asserted.

Dopamine transporter analysis SPECT

DaT-SPECT coregistered to a normative template, with the striatum parcellated into caudate and putamen subregions. Each returns a specific binding ratio, a z-score, a percentile and an asymmetry index against its contralateral pair — the measurements a reader would otherwise estimate by eye.

Patient   Normative template

Axial DaT-SPECT slices through the striatum with subregion volumes of interest overlaid, patient above and normative template below.

Patient above, age-matched template below, with the same subregion volumes overlaid on both. Binding ratios are computed inside those regions, not from a visual impression of the image.

Baseline and follow-up DaT-SPECT studies of the same patient acquired fourteen months apart.

The same patient, fourteen months later. Progression is quantified subregion by subregion rather than described.

Striatal binding ratio — illustrative case, age 68
SubregionSBR LZ LSBR RZ R
Ventral caudate0.81−3.420.81−2.48
Anterior caudate0.79−3.230.73−2.89
Anterior putamen1.11−2.430.81−3.85
Posterior putamen0.53−3.180.42−4.31

The posterior putamen is the most affected subregion and the right side more than the left — the rostro-caudal gradient and asymmetry that characterise a degenerative parkinsonian syndrome, here as numbers rather than as a reader's impression. Both caudate percentiles on the left sit at 0.00.

Progression at 14 months — same patient
SubregionBaseF/upΔ
Anterior caudate L0.790.45−54.8%
Anterior putamen L1.110.80−32.5%
Posterior putamen R0.420.29−36.6%
Ventral caudate L0.810.61−28.2%

Two studies, one year apart, on the same reconstruction pipeline and the same normative reference. Serial DaT imaging is only interpretable if the measurement is stable between timepoints — which is a software property, not a scanner one.

Epileptogenic focus localisation SISCOM

Ictal and interictal perfusion SPECT, subtracted from one another and coregistered to the patient's own MRI. The difference is thresholded, clustered, and mapped onto an anatomical atlas — so the candidate focus arrives with a location, a volume and a significance value rather than as a region a reader points at.

Axial MRI slices segmented by brain lobe, with the primary SISCOM focus outlined.

The primary focus outlined on the patient's own MRI, segmented by lobe — the anatomy a surgical team plans against.

SISCOM >2 SD   Ictal   Interictal

Three rows of matched axial slices: thresholded SISCOM clusters on MRI, ictal perfusion SPECT, and interictal perfusion SPECT.

Thresholded clusters above, with the ictal and interictal studies they were derived from beneath. The subtraction is auditable against its own inputs.

Primary focus

Inferiolateral parietal R

15.51 cm³

Most significant regions — illustrative case
RegionZExtent
Inferiolateral parietal R5.9113.15%
Cerebellum R5.7611.69%
Middle frontal gyrus R5.487.49%
Inferiolateral parietal L6.004.77%
Postcentral gyrus L6.002.85%

Significance alone would rank the left parietal and postcentral regions equal first. Extent — the proportion of the region actually occupied by the cluster — separates a large hyperperfused territory from a small intense one. Both numbers are needed, and the report gives both.

Coregistration quality
SPECT to MRI0.9686
MRI to reference0.9914
Ictal / interictal correlation0.99

Subtraction is only meaningful if the two studies are aligned. The registration metrics are reported alongside the result rather than assumed.

Amyloid burden on the Centiloid scale PET

Regional SUVR against a reference region, converted to Centiloid — the standardised scale on which 0 is a young healthy brain and 100 the typical Alzheimer's presentation. It is the unit anti-amyloid trials report in, and increasingly the unit eligibility is judged on, so the result travels between centres and between studies.

−20 12 29 150 POSITIVITY THRESHOLD −8.07
Illustrative case: −8.07 Centiloid, below the positivity threshold. The scale, not the picture, is what a second centre can act on.

Patient   Low-uptake reference   High-uptake reference

Three rows of axial amyloid PET slices: the patient, a low-uptake reference template, and a high-uptake reference template, with cortical regions of interest outlined.

The patient above, bracketed by low- and high-uptake reference templates with the same cortical regions outlined on all three. Grey–white differentiation is preserved in the top row and lost in the bottom — the visual signature the number encodes.

Regional SUVR — illustrative case, florbetaben
RegionSUVR LSUVR RZ R
Prefrontal0.961.00−0.58
Parietal1.021.150.45
Precuneus0.900.930.77
Cingulate0.810.82−0.41
Composite · cerebellar ref.SUVR 0.96−8.07 CL

Four reference regions are computed in parallel — cerebellum, pons, cerebellum plus brainstem, and cerebellar grey matter — because the choice of reference changes the answer, and a centre that uses a different one can still compare like with like.

Also available

Amyloid PET/MRI

Regions drawn on the patient's own MRI rather than a template, for cases where atrophy makes template-based parcellation unreliable. SUVR reported against the tracer-specific threshold.

neuroCloud is a quantification tool. Its output supports, and does not constitute, a diagnostic report. Cleared by ANVISA; FDA and other jurisdictions in progress — see the regulatory register above.

05

Deployment

The software goes into the environment you already run — PET, SPECT and MRI alike. No scanner replacement, no PACS migration, no change to how studies reach the reader.

No personal patient data leaves the site

The system does not receive identifiable patient data. That removes the largest single obstacle to information-governance sign-off, and it is a design decision rather than a configuration option.

Agnostic to what you already own

Independent of scanner OEM and model, file type, radionuclide and PACS. Installed equipment, including older systems, keeps producing reportable studies.

Integrated or alongside

Deploys into existing systems or runs alongside them, whichever suits the department. Clinical workflow is not redesigned around the software.

The original is never altered

Source studies remain intact and available for record and comparison throughout. Enhanced and original can be read side by side.

In-network or hosted

iPET and iPETcertum process inside the hospital network. neuroCloud is delivered as a hosted service, with data residency configured per deployment — so a centre without local compute still gets normative referencing and longitudinal comparison.

Normative data is maintained centrally

neuroCloud's population references and atlases are versioned and updated centrally, so every site quantifies against the same standard and results stay comparable between centres.

SITE BOUNDARY · NO IDENTIFIABLE PATIENT DATA LEAVES PET / PET-CT SPECT MRI iPET · iPETcertum neuroCloud PROCESSING & QUANTIFICATION Enhanced study ACCELERATED ACQUISITION Quantification SUV · VOLUME · Z-SCORE Original study RETAINED, UNALTERED DICOM PACS workstation Agnostic to scanner OEM and model · file type · radionuclide · PACS
PET, SPECT and MRI studies enter as DICOM and return as enhanced images, quantification and normative reports. Studies are handled without identifiable patient data, and the source study is preserved throughout.

HIPAA · GDPR · PDPA compliant across all supported regions.

06

For radiopharmaceutical developers

ONE LESION · THREE TIMEPOINTS · IDENTICAL CRITERIA Baseline 146.1 ml · SUVmax 16.5 Cycle 2 61.4 ml · SUVmax 9.4 Cycle 4 17.8 ml · SUVmax 3.8 Illustrative. Response is measured, not described, because the criteria did not move between scans.

A tracer produces an image. What a sponsor actually needs is a number — comparable between sites, stable between readers, and defensible to a regulator.

That quantification layer is the part of a launch most often assembled late, per product, by a different vendor each time. We build it once, across a portfolio, so a sponsor's prostate, renal and neuro assets report on one standard rather than three.

We work in trials, in expanded access, and at commercial launch. The software is agnostic to isotope and to scanner manufacturer, which matters when a product's distribution reaches sites that have never run that tracer before.

For theranostic pairs, the same criteria applied across successive scans measure progression rather than describing it — lesion volume, SUV and SUVmax compared before and after treatment on a consistent basis.

How an engagement runs

  1. Feasibility on your data

    A defined dataset, evaluated against the clinical task your label actually claims — not a generic performance benchmark.

  2. Reader and site variability

    Where the variance is: acquisition, reconstruction, reader, or something else. The residual is usually the interesting part.

  3. Validation and evidence

    Assembled to carry a submission, with the run manifest, software version and operator recorded against every result.

  4. Deployment at launch

    Integration into the sites reading your tracer, under one licence across the portfolio rather than one per product.

07

Why the number holds

Most image AI asks to be trusted. Ours asks to be checked. In a field where a generative model can add a lesion that was never scanned, the discipline below is the product.

The clinician sets the criteria

Segmentation runs against minimum and maximum SUV and volume parameters defined by the reader, not against a model's internal opinion of what counts as a lesion. The decision rule is visible and adjustable, which is what makes a result contestable — and therefore usable.

Voxel-level, not impression-level

Values derive from uptake at voxel level. Two readers applying the same criteria to the same study obtain the same figures. Subjectivity is removed from the measurement without being removed from the diagnosis.

Nothing invented

Enhancement suppresses noise and brings out detail the acquisition already captured. It does not synthesise structure that was not there. Fine lesions are retained, artefacts are not introduced, and the original is preserved alongside the enhanced image so a reader can check the claim rather than accept it. Fidelity is reported against RMSE and SNR.

Versioned and re-executable

Every result carries the software version, the parameters applied and the operator. A finding from eighteen months ago can be reproduced exactly, which is the difference between evidence and output.

Vendor-neutral by construction

Brand-agnostic and isotope-agnostic, integrating into existing scanner and PACS environments. A multi-site study does not need matched hardware to produce comparable numbers.

Built under a quality system

ISO 13485 certified, with clearances obtained and maintained per product and per jurisdiction rather than claimed in aggregate.

Where generative enhancement runs into trouble

A growing number of tools improve a scan by generating a cleaner version of it. The images look excellent. The difficulty is that a model trained to produce plausible images will produce a plausible image whether or not the underlying signal supports it — and the reader has no way to tell the two cases apart.

The generative problem

  • Structure can be synthesised that the scanner never recorded. A lesion can be added; a faint one can be smoothed away. Both outputs look correct.
  • The same study does not reliably yield the same image. Change the model version, or present a scanner, tracer or body habitus outside the training distribution, and the output shifts — silently, with no flag to the reader.
  • Quantification moves with the image. Values read off a generated reconstruction belong to the model, not to the acquisition.
  • That breaks theranostics outright, where the entire question is whether this timepoint differs from the last.
  • Validation is difficult to sustain: retraining changes behaviour, so the evidence supporting clearance ages with every model update.

How Claritas behaves

  • Deterministic. The same study with the same parameters returns the same result, today and in two years.
  • Nothing is synthesised. Noise is suppressed and detail already captured is brought out. Fine lesions are retained and no artefacts are introduced.
  • SUVmin preserved to within 1% of the original; SUV statistics reported separately rather than as one figure.
  • The original is never altered and sits alongside the enhanced study, so the claim can be checked rather than taken on trust.
  • Fidelity is measured against the source by RMSE and SNR — not against a subjective judgement of how good the picture looks.

This is why the same engine can carry both enhancement and quantification. If enhancement altered the measurement, iPETcertum could not be run on an enhanced scan and the two products would have to stay apart. Because it does not, an accelerated, reduced-dose acquisition still yields a number a physician can report and a sponsor can submit.

Bring us a dataset.

The fastest way to evaluate this is to run it on studies you already have and compare the output against your current read. We will tell you where it does not help, which is usually the more useful half of the answer.

Request an evaluation Request the regulatory annex